Part Number Hot Search : 
VPX3226 00RL7 24C32 MB86291 2SD1409 0XW12 AX88178 HBU150
Product Description
Full Text Search
 

To Download LPS45-M Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
 Embedded Power for Business-Critical Continuity
LPS40-M Series
40 Watts Medical
Total Power: Input Voltage: # of Outputs: 40-55 Watts 85-264 VAC 120-300 VDC Triple
Rev. 04.14.06 LPS40-M Series 1 of 3
RoHS
Electrical Specifications
Special Features
* * * * * * * * * * * Medical safety approvals Universal input 3" x 5" footprint Less than 1U high Remote sense on main output Low output ripple Adjustable 5V output Overvoltage protection Overload protection 110 KHz switching frequency LPX40 enclosure kit available
Input
Input range Frequency Inrush current Input current Efficiency EMI filter Safety ground leakage current 85-264 VAC; 120-300 VDC 47-440 Hz <18A peak @ 115VAC; <36A peak @ 230VAC, cold start @ 25C 1A max. (RMS) @ 115VAC 70% typical at full load FCC Class B conducted; CISPR 22 Class A conducted; EN55022 Class A conducted; VDE 0878 PT3 Class A conducted <75A @ 50/60 Hz, 264 VAC input
Output
Maximum power Adjustment range Hold-up time Overload protection 40W for convection; 55W with 30CFM forced air -5, +10% minimum 20ms @ 40W load, 115VAC nominal line
Safety
0750/EN60601-1 (IEC601) UL2601 E182560 CSA 22.2 No.601.1 LR109492 CE Mark (LVD) VDE UL CSA
Short circuit protection on all outputs Case overload protected @ 110 -145% above peak rating Overvoltage protection 5V output; 5.7 to 6.7VDC. Other outputs 10% to 25% above nominal output Remote sense Compensates for 0.5 V lead drop minimum. Will operate without remote sense connected. Reverse connection protected.
Embedded Power for Business-Critical Continuity
Rev. 04.14.06 LPS40-M Series 2 of 3
Environmental Specifications
Operating temperature: 0 to 50C ambient. Derate each output 2.5% per degree from 50 to 70C Storage temperature: -40C to +85C Temperature coefficient: 0.4% per C Electromagnetic susceptibility: Designed to meet IEC 801, -2, -3, -4, -5, -6, Level 3 Humidity: Operating; non-condensing 5% to 95% RH Vibration: MTBF demonstrated: Three orthogonal axes, sweep at 1 oct/min, 5 min. dwell at four major resonances 0.75G peak 5Hz to 500Hz, operational >550,000 hours at full load and 25C ambient conditions
Ordering Information
Maximum Load Model Output Minimum Maximum Load with with 30CFM Number Voltage Load Convection Cooling forced Air LPS42-M 5V 0A 8A 11A LPS43-M 12V LPS44-M 15V LPS45-M 24V 0A 0A 0A 3.3A 2.6A 1.6A 4.5A 3.6A 2.3A Ripple P/P Peak Load 1 Regulation 2 (PARD) 3 12A 2% 50 mV 5A 4A 2.5A 2% 2% 2% 120 mV 150mV 240mV
1. Peak current lasting <30 seconds with a maximum 10% duty cycle. 2. At 25C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings. 3. Peak-to-peak with 20 MHz bandwidth and 10 F in parallel with a 0.1 F capacitor at rated line voltage and load ranges.
Power Derating Curve
55W 30 CFM Forced Air Cooling 40W
30W Output Power (Watts) 20W
Natural Convection Cooling
27.5W 20W
10W 0W 0C 10C 20C 30C 40C 50C 60C 70C TA - Ambient Temperature - C
Embedded Power for Business-Critical Continuity
Mechanical Drawing Astec Power
0.22" (5.6) M1 GND
SK201
Rev. 04.14.06 LPS40-M Series 3 of 3
4.55" (115.6)
5.00" (127)
M2
_
+
SK 2
3.00" (76.2) 2.55" (64.8)
SK 1
Remote Sense Connector Voltage Adj. Pot
5810 Van Allen Way Carlsbad, CA 92008 USA Telephone: +1 760 930 4600 Facsimile: +1 760 930 0698 Technical Support: +1 888 41 ASTEC or +1 407 241 2752 Waterfront Business Park Merry Hill, Dudley West Midlands, DY5 1LX United Kingdom Telephone: +44 (0) 1384 842 211 Facsimile: +44 (0) 1384 843 355 Units 2111-2116, Level 21 Tower 1, Metroplaza 223, Hing Fong Road Kwai Fong, New Territories Hong Kong Telephone: +852 2437 9662 Facsimile: +852 2402 4426
3 1
N L
1 2 3 4 5 6
0.15" (4) DIA (4 places)
For global contact, visit:
1.14" (29)
www.astecpower.com technicalsupport@astec.com
While every precaution has been taken to ensure accuracy and completeness in this literature, Astec Power assumes no responsibility, and disclaims all liability for damages resulting from use of this information or for any errors or omissions. Printed in US A
0.12" (3)
Pin Assignments
Connector LPS42-M LPS43-M LPS44-M LPS45-M SK1-1 SK1-3 SK2-1 SK2- 2 SK2- 3 SK2-4 SK2-5 SK2-6 SK201-1 SK201-2 Neutral Line +5V +5V +5V Common Common +Sense -Sense Neutral Line +12V +12V +12V Common Common +Sense -Sense Neutral Line +15V +15V +15V Common Common +Sense -Sense Neutral Line +24V +24V +24V Common Common +Sense -Sense
Mating Connectors
AC Input: Molex 09-50-8031 (USA) 09-91-0300 (UK) PINS: 08-58-0111
Emerson Network Power. The global leader in enabling business-critical continuity. AC Power Connectivity DC Power Embedded Power Inbound Power Integrated Cabinet Solutions Outside Plant Precision Cooling Site Monitoring and Services EmersonNetworkPower. com
Emerson Network Power and the Emerson Network Power logo are trademarks and service marks of Emerson Electric Co. (c)2006 Emerson Electric Co.
Common Common Common Common
Molex 09-50-8061 (USA) Not required for (-T) option 09-91-0600 (UK) PINS: 08-58-0111 Remote Sense: Molex 22-01-2025 PINS: 08-50-0114 Astec Connector Kit #70-841-006, includes all of the above 1. Specifications subject to change without notice. 2. All dimensions in inches (mm), tolerance is 0.02" (0.5mm) 3. 4. 5. 6. 7. Mounting holes M1 and M2 should be grounded for EMI purposes. Mounting hole M1 is safety ground connection. Specifications are for convection rating at factory settings at 115 VAC input, 25C unless otherwise stated. Warranty: 1 year Weight: 0.5lbs/0.23kg
DC Outputs:


▲Up To Search▲   

 
Price & Availability of LPS45-M

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .